Research on Phase Change Characteristic of Miniature Gallium Fixed-point Using Quasi-adiabatic Measurement Method |
XU Chun-yuan1,2,XIA Cai-juan1,HAO Xiao-peng2,SUN Jian-ping2,WEN Ping2,SONG Jian2,YUAN Zun-dong2 |
1. School of Science, Xi'an Polytechnic University, Xi’an, Shaanxi 710048, China
2. National Institute of Metrology, Beijing 100029, China |
|
|
Abstract The formation process of miniature Ga fixed-point which can be used for in situ calibration, the structure and working principle of the quasi-adiabatic testing system are described.The phase change characteristic of miniature Ga fixed-point is measured using this quasi-adiabatic testing system. The results show that the melting point of Ga fixed-point has strong linear relationship with the heating power. The linear fitting factor R2 is 0.995 7. The lower the heating poweris, the closer to the standard temperature of Ga fixed-point it becomes. Further more, the repeatability of the melting points of Ga is better than 2 mK under the same quasi-adiabatic testing system.
|
Received: 05 April 2016
Published: 28 December 2016
|
|
Corresponding Authors:
Xiaopeng Hao
E-mail: haoxp@nim.ac.cn
|
|
|
|
[1]Thomas H, Quinn T J. The International Temperature Scale of 1990 (ITS-90)[J]. Ibid,1990,27(1):3-10.
[2]郝小鹏,原遵东,王景辉,等.1.6微米近红外辐射温度计研制[C]//2010中国仪器仪表与测控技术大会,深圳,2010.
[3]赵婉君,郝小鹏,原遵东,等. 900 nm照度辐射温度计的研制[J]. 计量学报,2011,32(4):324-328.
[4]黄燊彦,郝小鹏,王景辉,等.铯、钠热管标准黑体辐射源研制及性能评价[J]. 计量学报,2014,35(2):120-124.
[5]Hao X P,Mcevoy H,Machin G, et al. A comparison of the In, Sn,Zn and Al fixed points by radiation thermometry between NIM and NPL and verification of the NPL blackbody reference sources from 156 ℃ to 1000 ℃[J]. Measurement Science and Technology,2013,24(7):1-6.
[6]Hao X P,Yuan Z D, Huang S Y, et al. Study on the infrared lens-free irradiation thermometer based on InGaAs detector at NIM[J]. International Journal of Thermophysics,2015,36(12):1-10.
[7]郝小鹏,宋健,孙建平,等.风云卫星的红外遥感亮度温度国家计量标准装置[J].光学精密工程,2015, 23(7):1845-1851.
[8]孙建平.准绝热密封氩三相点复现装置[J]. 计量学报,2011,32(6):481-484.
[9]Ohring G, Tansock J, Emery W, et al. Achieving satellite instrument calibration for climate change[J]. Eos Transactions American Geophysical Union,2007,88(11):136-136.
[10]Best F A, Revercomb H E. On-orbit Absolute Temperature Calibration Using Multiple Phase Change Materials: Overview of Recent Technology Advancements[J]. Proc Spie,2010,7857:253-257.
[11]Burdakin A,Khlevnoy B,Samoylov M, et al. Melting points of gallium and of binary eutectics with gallium realized in small cells[J]. Metrologia,2008,45(1):75-82.
[12]曾凡超,曾兵,孙建平,等.基于微型Ga-Sn共晶点的精密铂电阻温度计现场标定[J]. 计量学报,2015,36(6A):28-31.