A Method for Measurement of the Inner Volume of Bell Prover |
LIU Yi-ping1,SONG Jin1,XU Yao1,QUAN Xiao-jun2 |
1.Shanghai Institute of Measurement and Testing Technology, Shanghai 201203, China
2.Shanghai Jiao Tong University, Shanghai 200240, China |
|
|
Abstract The inner volume of a 1000L bell prover is measured by a new facility. A chord length measuring element is made up of two length gauges and a rotating arm. The whole volume of the bell is measured off 256 cross sections with axially equal spacing, and the spatial coordinates of the points, at which length gauges contact the bell’s inner wall,are obtained with a sampling interval of 15° to reproduce every section profile. A new sweep method is introduced to calculate the areas of section profiles. The comparison with the method for the assessment of departure from roundness shows that the results of the two methods coincide with each other when small sampling interval is adopted. A data sheet for all element volumes along bell axial is presented as calibration result.
|
Received: 15 November 2014
Published: 14 October 2016
|
|
|
|
|
[1]王池. 气体流量量值溯源体系探讨[J]. 中国计量学院学报, 2003, 14(3): 170-173.
[2]国家质量监督检验检疫总局. JJG 165-2005, 钟罩式气体流量标准装置 [S]. 2005.
[3]Brunnneman H, Aschenbrenner A. Volume with guarantee[R]. Physikalisch-Technische Bundesanstalt, 1983.
[4]王池, 王东伟, 徐英华.钟罩式气体流量标准装置不确定度分析[J]. 计量学报, 2001, 22(3): 219-222.
[5]王池, 崔骊水. 气体流量标准装置钟罩外径测量方法的研究[J]. 计量学报, 2007, 28(3): 239-242.
[6]崔骊水. 钟罩检定方法的研究[D]. 北京:中国计量科学研究院, 2005.
[7]Cui L S, Li X. A kind of New Calibration Method for the Volume of Bell Prover[C] // Flomeko. The 15th International Flow Measurement Conference. Taipei, 2010, 82-86.
[8]李旭, 崔骊水, 孟涛, 等. 基于尺寸法的标定钟罩式气体流量标准装置的新方法[J]. 计量技术, 2008, (9): 42-45.
[9]国家质量监督检验检疫总局. GB/T 7234—2004, 产品几何量技术规范(GPS)圆度测量 术语、定义及参数[S]. 2004. |
|
|
|