A Combination of Measuring Methods for Modulation Transfer Characteristic of Proximity X-ray Imaging System |
MU Yi-ning,WANG He,LI Ye,LI Ping,ZHU Ling-xi |
Changchun University of Science and Technology, Changchun, Jilin 130022, China |
|
|
Abstract Using the evaluating principle of modulation transfer function, an optical modulation transfer characteristic detection methodology of proximity focusing X-ray imaging system is brought up on the basis of the signal transfer relationship between each imaging subsystem. From the constraints of image-intensifier tubes, the generation process of ray source and imaging restriction of its spot size constraints, the constraints of noise three perspectives to calibrate theoretical modulation transfer characteristic of proximity X-ray imaging system. Change CCD telescopic system of proximity X-ray imaging system back end to CCD microscopy system, to multi-spatial resolution balls of lead grating imaging and MTF curve fitting, and as the experimental results suggest, the curve fitting precision can be significantly improved by time domain equalization. Fitting results indicate that modulation transfer function still are equal to 10% when spatial resolution reaches 10lp/mm, this is consistent with that of the theoretical calibration. So that the resolution of traditional proximity X-ray imaging system is really the bottleneck.
|
Received: 29 October 2014
Published: 29 July 2016
|
|
|
|
|
[1]赵宝昇,王治平.X射线像增强器象质评价[J].高速摄影与光子学, 1991, 20(3):292-299.
[2]潘俊杰, 赵宝升, 赛小锋, 等. 影响X 射线像增强器分辨率的因素分析[J]. 光子学报, 2008, 37(6): 1116-1118.
[3]李伟, 赵宝升, 张兴华, 等. 双近贴式X 射线像增强器成像系统的三维噪音测量及其分析技术[J]. 光子学报, 2009, 38(8): 1932-1936.
[4]李伟, 赵宝升, 赵菲菲, 等. 双近贴式X 射线像增强器成像不均匀性的分析与校正[J]. 光子学报, 2009, 38(6): 1353-1357.
[5]王凯歌, 牛憨笨, 郭金川. 微束斑X射线源聚焦系统的数值研究[J]. 光子学报, 1999, 28(12): 1141-1145.
[6]王凯歌, 牛憨笨, 万云, 等. 微束斑X射线源的边界元法分析[J]. 光子学报, 2004, 33(8): 228-232.
[7]Dhez P,Chevallier P, Lucatorto T B, et al. Instrumental aspects of x-ray microbeams in the range above 1 keV[J]. Rev Sci Instrum, 1999, 70(4), 1907-1920.
[8]Khacef A, Viladrosa R, Cachoncinlle C, et al. High repetition rate compact source of nanosecond pulses of 5-100 keV X-ray photos [J]. Rev Sci Instrum, 1997,68(6): 2292-2297.
[9]王桂英,黄柳红,沈建国. 利用CCD 测定X 射线增感屏的调制传递函数[J]. 计量学报,1997,18 (3):187-191.
[10]吴金杰,王培玮,段小娟,等. 低能X射线基准辐射装置的建立[J]. 计量学报,2011,32(6A):9-13.
[11]潘京生.微通道板及其主要特征性能[J].应用光学,2004,25(5):25-29.
[12]李野. 便携式低强度X射线实时成像系统关键技术研究[D]. 长春:长春理工大学,2010.
[13]母一宁,于林韬, 朱一峰,等. 基于增量式光电编码器细分的自适应滤波技术研究[J]. 计量学报, 2012, 33(2): 110-114. |
|
|
|