Development and Measurement of Relative Humidity Parameter Calibration Device in Negative Temperature Environment |
GONG Juan,LÜ Guo-yi,HE Meng |
AVIC Changcheng Institute of Metrology and Measurement, Beijing 100095, China |
|
|
Abstract A negative temperature environment relative humidity generator developed by the Changcheng Institute of Metrology & Measurement is introduced, which is based on double temperature and double pressure method. By setting the required temperature and relative humidity of environment, to calculate the value of the corresponding dew point, automatically set full and internal pressure and the temperature of the device and measuring chamber, thereby generating set temperature under the condition of the relative humidity of the sample gas, the relative humidity range is 10%RH~95%RH (-60~40 ℃). The relative humidity parameter calibration process and the corresponding experimental measurement results are analyzed.
|
|
|
|
|
|