Research on the Temperature Standard for the Thermometer Sensor Pair of Heat Meter |
QIU Ping,XIE Li,WANG chi |
National Institute of Metrology, Beijing 100013 |
|
|
Abstract Temperature standard instruments are studied for the temperature sensor pair of heat meter. Two grade B industrial platinum resistance thermometers (IPRTs) are tested. The experiments showed that the stability of high precise IPRT is 3 mk in the temperature range from 0 ℃ to 90 ℃ per year.The resistance drifts of the IPRTs were eliminated by using W value calculation and the accuracy of the calibration is improved.
|
|
|
|
|
|
[1]国家质量监督检验检疫总局,JJG 225-2001 热能表检定规程[S].2001
[2]国家质量监督检验检疫总局,JJG 229-2010 工业铂、铜热电阻检定规程[S].2011. |
|
|
|