Activity Measurement of 18F with TDCR-LS Methods |
LIU Hao-ran1,LIU Jia-cheng2,LIANG Jun-cheng2,YANG Yuan-di2,YUAN Da-qing1 |
1.China Institute of Atomic Energy, Beijing 102413, China
2.National Institute of Metrology, Beijing 100013, China |
|
|
Abstract As an important short half-life radionuclide,the 18F is widely used in nuclear medicine for positron emission tomography(PET).It disintegrates to 18O by β+ emission(96.86%)and electron capture(3.14%).Since the very high detection efficiency of β+ decay branch in the liquid scintillation counter,the radioactive solution of 18F has been standardized by TDCR-LS method.The results indicate that the detection efficiency of β+ emission in Ultima GoldTM AB cocktail is close to 1 and the uncertainty of 18F by TDCR-LS method is equal to 0.6%(k=2).
|
|
|
|
|
|
[1]DDEP-Decay Data Evaluation Project[DB],http://www.nucleide.org/DDEP_WG/Nuclides/F-18_tables.
[2]Broda R,Pochwalski K.The enhanced triple to double coincidence ratio(ETDCR)method for standardization of radio-nuclides by liquid scintillation counting[J].Nucl Instrum Methods,1992,A312:85-89.
[3]Broda R,Cassette P,Kossert K.Radionuclide metrology using liquid scintillation counting[J].Metrologia,2007,44:36-52.
[4]Grau Carles A.New Methods for the Determination of beta Spectra Shapefactor Coefficients[J],Appl Radiat Isot,1995,46(2):125-128.
[5]Birks J B.The Theory and Practice of Scintillation Counting[M],London:Pergamon Press,1960,186-187.
[6]吴永乐,液闪TDCR方法测量氚水活度[D].北京:中国原子能科学研究院, 2010.
[7]Broda R,Maletka K,et al.Study of the influence of the LS-cocktail composition for the standardization of radionuclides using the TDCR mode1[J].Appl Radiat Isot,2002,56:285-289.
[8]Miguel R,Eduardo G T,Leonor R B.Standardization of 18F by coincidence and LSC methods[J].Appl Radiat Isot,2006,64:1192-1202.
[9]梁珺成,应用CIEMAT/NIST方法测量轨道电子俘获核素活度的研究[D].北京:中国计量科学研究院,2008.
[10]Bouchard J,Cassette P.MAC3:an electronic module for the processing of pulses delivered by a three photomultiplier liquid scintillation counting system[J].Appl.Radiat.Isot.2000,52:669-672.
|
|
|
|