Acta Metrologica Sinica  2011, Vol. 32 Issue (z1): 18-21    DOI: 10.3969/j.issn.1000-1158.2011.z1.05
Current Issue | Archive | Adv Search |
Calibration Method of γ-ray Reference Radiation Field with Low Air Kerma Rate
LI De-hong1,HUANG Chen-wei2,WU Hui-li1,LI Xing-dong1,WU Meng-meng1,3
1. National Institute of Metrology, Beijing 100013, China
2. Henan Province Institute of Metrology, Zhengzhou, Henan 450008, China
3. Chengdu University of Technology, Chengdu, Sichuan 610051, China
Download: PDF (639 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      
Abstract  In order to verify whether the use of large volume ionization chamber can measure the low rate of reference air kerma radiation dose rate,a calibrated high pressure ionization chamber and atmospheric pressure ionization chamber were used to measure in reference radiation field produced by different activity Cs-137 source, the difference between the two measurements is less than 0.8%, the difference between the calculated values obtained by quadratic approximation formula and measured values is less than 0.3%. In addition, through verifying the radiation field by the inverse square law,the scattering share of the radiation field was measured by the shadow mask method,and the results show that the radiation field can meet the requirements of ISO-4037.
Key wordsMetrology      Air kerma      Calibration      Ionization chamber      Scattered radiation      Shadow shield     
PACS:  TB98  
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
LI De-hong
HUANG Chen-wei
WU Hui-li
LI Xing-dong
WU Meng-meng
Cite this article:   
LI De-hong,HUANG Chen-wei,WU Hui-li, et al. Calibration Method of γ-ray Reference Radiation Field with Low Air Kerma Rate[J]. Acta Metrologica Sinica, 2011, 32(z1): 18-21.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2011.z1.05     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2011/V32/Iz1/18
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech