Abstract  Auto-calibration system of vector hydrophone has very important value in acoustic measurement. The system can accomplish low frequency auto-calibration of vector hydrophone. It can measure the signal amplitude of the standard hydrophone by dynamic signal acquisition. The sound field distribution is tested in the standing wave field. The standard uncertainty and expanded uncertainty of the system are analyzed and discussed. The comparison testing is given for pressure sensitivity. The expanded uncertainty of the system is 1.6dB.
Key wordsMetrology      Vector hydrophone      Auto-calibration      Standing wave field      Uncertainty analysis     
PACS:  TB95  
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
LIU Jin-chang
FEI Teng
LIU Ming-bo
Liao Shu-chang
Cite this article:   
LIU Jin-chang,FEI Teng,LIU Ming-bo, et al. Study and Implementation of the Auto-calibration System of Vector Hydrophone[J]. Acta Metrologica Sinica, 2011, 32(6): 526-530.
URL:  
http://jlxb.china-csm.org:81/Jwk_jlxb/EN/10.3969/j.issn.1000-1158.2011.06.10     OR     http://jlxb.china-csm.org:81/Jwk_jlxb/EN/Y2011/V32/I6/526
Copyright © Editorial Board of Acta Metrologica Sinica
Supported by:Beijing Magtech