Discussion on the Calculation Method for Reference Value and Degrees of Equivalence in Comparisons |
YUAN Zun-dong,WANG Chi,LIN Yan-dong |
National Institute of Metrology, Beijing 100013, China |
|
|
Abstract Based on considering the correlation between the laboratory result and reference value, the weighted average method for calculating the reference value and the method for calculating the uncertainty of the degrees of equivalence are introduced. Furthermore,the uncertainty introduced in the comparison and correction process contributes to the weights calculating the weighted average method are presented. It is pointed out that the influence of correlation of the laboratory measured values on analyzing and calculating the uncertainty of the reference value or that of optimal estimated value of the external independent values should be considered.
|
|
|
|
|
|
[1]CIPM.Guidelines for CIPM key comparisons [EB].http://www.bipm.org/utils/en/pdf/guidelines.pdf,2009-06-05.
[2]Cox M G. A discussion of approaches for determining a reference value in the analysis of key-comparison data[R]. NPL Report CISE 42/99,1999.
[3]Cox M G. The evaluation of key comparison data [J]. Metrologia, 2002, 39(6):589~595.
[4]Elster C, Link A. Analysis of key comparison data: assessment of current methods for determining a reference value [J]. Meas Sci Technol, 2001, 12(9):1431~1438.
[5]Steele A G, Wood B M, Douglas R J. Exclusive statistics: simple treatment of the unavoidable correlations from key comparison reference values [J]. Metrologia, 2001, 38(6): 483~488.
[6]Douglas R J, Steele A G. Avoiding a complex reference value [J]. Metrologia, 2008, 45(3):L13~L15.
[7]刘智敏. 关键比对的等效度评定[J], 中国计量学院学报, 2007, 18(1):6~13.
[8]原遵东, 王池, 林延东. 确定量值比对参考值的准则[J]. 仪器仪表学报,2010,31(10):2381-2385. |
|
|
|