|
|
The Experiment Study for SSE on Calibration of Infrared Thermometer under Lower Temperature Condition |
ZHAO Quan-qin1,LI Wen-jun1,SUN Jian1,YU Shi-fan2,ZHANG Long-fei1,CUI Chao2 |
1.China Jiliang University, Hangzhou, Zhejiang 310018, China
2. Zhejiang Province Institute of Metrology, Hangzhou, Zhejiang 310013, China |
|
|
Abstract The phenomenon of size-of-source effect (SSE) is discussed and the coefficient of SSE is used for estimating the SSE property of the infrared thermometer. The experiment apparatus used for measuring SSE under lower temperatures condition is described, and several types of infrared thermometers are applied to the measurements of SSE. The experiment results indicate that the influence of SSE may be not considered at the diameters of the diaphragm hole being more than 40mm. This conclusion is suitable to the most industrial infrared thermometers.
|
|
|
|
|
|
[1]段宁宇,赵琪,原遵东,等.辐射源尺寸效应研究[J].计量学报,1996,17(3): 161-166.
[2]原遵东.消除背景辐射影响的辐射源尺寸效应测量模型[J].光学学报, 2011, 31(12): 101-106.
[3]郭莉莉,原遵东,邢波,等.工业辐射温度计的辐射源尺寸效应(SSE)测量[J].计量技术,2010,(9):6-9.
[4]Saunders P, Edgar H. On the characterization and correction of the size-of-source effect in radiation thermometers[J]. Metrologia,2009, 46:62-74.
[5]Bloembergen P. Analytical represtations of the Size-of-Source Effect[J]. Metrologia, 2009, 46(5):534-543.
[6]Bloembergen P. On the correction for the size-of-source effect corrupted by background radiation[C]//Proceedings of TEMPMEKO 1999,the 7th International Symosiun on Temperature and Thermal Measurements in Industry and Science, Delft, 1999, 607-612. |
|
|
|