Research on the Calibration Method of Reference Solar Cell |
ZHANG Jun-chao,XIONG Li-min,MENG Hai-feng,HE Ying-wei,ZHANG Bi-feng |
National Institute of Metrology, Beijing 100029, China |
|
|
Abstract Based on differential spectral responsivity method, a calibration equipment for reference solar cell is established. The absolute responsivity of reference cell varied with temperature of cell and irradiance intensity of bias light are analyzed. The relationship between the short circuit current and the bias current of the solar cell is obtained by numerical simulation techniques. The uncertainty of the result is 1.2% (k=2). Based on outdoor method, a reference cell calibration equipment is established. A detailed uncertainty evaluation of the result is carried out, the value is 2.38% (k=2). Through numerical calculation and analysis of the En value (0.95), it is found that the results measured by the two methods are equivalent.
|
Received: 13 May 2015
Published: 28 February 2017
|
|
Corresponding Authors:
Jun-chao ZHANG
E-mail: zhangjc@nim.ac.cn
|
|
|
|
[1]孙皓,熊利民.不同类型太阳电池光电参数的测量方法[J].中国计量,2011,(4):87-89.
[2]孙皓,熊利民.光谱失配对标定太阳电池产生的影响分析[J].计量学报,2011,32(5):432-436.
[3]Hartman J S, Lind M A. Spectral response measurements for solar cells[J]. Sol Cells, 1982, 7(1-2):147-157.
[4]Boivin L P, Budde W, Dodd C X, et al. Spectral response measurement apparatus for large area solar cells[J]. Applied Optics,1986, 25(16):2715-2719.
[5]Metzdorf J. Calibration of solar cells. 1: the differential spectral responsivity method [J]. Applied Optics, 1987,26(9): 1701–1708.
[6]Metzdorf J, Winter S, Wittchen T. Radiometry in photovoltaics: calibration of reference solar cells and evaluation of reference values [J]. Metrologia, 2000, 37(5): 573–578.
[7]Müllejans H, Ioannides A, Kenny R, et al. Spectral mismatch in calibration of photovoltaic reference devices by global sunlight method [J]. Measurement Science Abd Technology, 2005, 16(6):1250-1254.
[8]Müllejans H, Zaaiman W, Dunlop E D, et al. Calibration of photovoltaic reference cells by the global sunlight method[J]. Metrologia, 2005,42(5):360-367.
[9]张亚,袁亚飞,汤浩军.太阳电池空间标定技术初探[J]. 宇航计测技术,2010,30(1):28-32.
[10]刘海涛,桑识宇,翟永辉,等.标准光伏电池在拉萨自然阳光下的一级标定[J]. 中国测试,2012,38(1):28-31.
[11]Shimokawa R, Nagamine F, Miyake Y, et al. Japanese indoor calibration method for the reference solar cell and comparison with outdoor calibration [J]. Japanese Journal of Applied Physics, 1987, 26(1): 86–91.
[12]Emery K A, Osterwald C R, Rummel S, et al. A Comparison of Photovoltaic Calibration Methods[C]//Proc 9th European Photovoltaic Solar Energy Conf, Freiburg, W. Germany, 1989. |
|
|
|