|
|
APMP Pilot Comparison of Low-g Shock Acceleration Primary Calibration |
SUN Qiao,HU Hong-bo,BAI Jie |
National Institute of Metrology, Beijing 100029, China |
|
|
Abstract As pilot lab, National Institute of Metrology, organized Asia Pacific Metrology Programme pilot comparison of primary low-g shock acceleration calibration from August, 2013 to December, 2014. The comparison covers primary shock calibration systems using laser interferometry with two basic types of low-g shock exciters from 4 participating labs. The peak value of shock acceleration for comparison of monopole shock pulse is from 500m/s2 to 5000m/s2 with pulse duration from 0.5ms to 5ms, and for comparison of dipole shock pulse is 1000m/s2 with pulse duration from 0.03ms to 0.2ms. The comparison results show that the deviations of participants’ measurement results from reference values are within the associated uncertainties.
|
Received: 10 April 2015
Published: 28 February 2017
|
|
|
|
|
[1]中国计量测试学会.一级注册计量师基础知识及专业务实[M].北京:中国计量出版社,2011.
[2]ISO 2001 International Standard 16063-13 Methods for the calibration of vibration and shock transducers—Part 13: Primary shock calibration using laser interferometry[S]. Geneva: International Organization for Standardization, 2001.
[3]Link A,von Martens H J. Calibration of accelerometers by shock excitation and laser interferometry[J].Shock & Vibration, 2000, 7(2):101-112.
[4]Nozato H, Usuda T, Oota A, et al. Calibration of vibration pick-ups with laser interferometry: part IV Development of a shock acceleration exciter and calibration system[J]. Meas Sci Technol,2010, 21(6):65107-65116.
[5]Sun Q,Wang J L,Hu H B . A primary standard for low-g shock calibration by laser interferometry[J].Meas Sci Technol , 2014, 25(7): 075003.
[6]孙桥,王建林,胡红波,等.低g值冲击加速度的激光绝对法校准 [J]. 计量学报, 2015, 36(2): 145-148.
[7]Huang Y C,Chen J K,Ho H C, et al. The set up of primary calibration system for shock acceleration in NML[J].Measurement, 2012,45(10):2383–2387.
[8]SPEKTRA. Data sheet of SE-220 HOP-MS Hopkinson Bar Shock Exciter [ED]. http://www.spektra-dresden.com/products/vibration-and-shok-exciter/shock-exciter.html.
[9]Sun Q, Hu H B. Final Report on Pilot Comparison of Low Intensity Shock APMP.AUV.V-P1 [J]. Metrologia Tech Suppl, 2015, 52: 075003. |
|
|
|