Development of the Fourth-generation National Primary Standard Apparatus of Spectral Radiometry and Color Temperature |
DAI Cai-hong,WU Zhi-feng,OUYANG Hui-quan,HUANG Bo,YU Jia-lin |
Division of Metrology in Optics and Laser, National Institute of Metrology, Beijing 100013, China |
|
|
Abstract Based on the new high temperature blackbody BB3500M and optical feedback control system, the fourth-generation national primary standard apparatus of spectral radiometry and color temperature are evaluated and established. The stability, uniformity, repeatability and measurement uncertainty of the apparatus are improved greatly. The measurement uncertainty of temperature was decreased to 0.64 K (k=1) when BB3500M blackbody is operating at 2 980 K, which is traceable to the Pt-C and Re-C fixed point blackbodies of NIM. Wavelength measurement range of spectral radiance and spectral irradiance are extended to 220~2 550 nm and 230~2 550 nm respectively, which meets the full wavelength requirements of CCPR-S1 spectral radiance international comparison. The measurement uncertainty of color temperature and distribution temperature is promoted to 1.6 K at 2 353 K and 2.1 K at 2 856 K
|
|
|
|
|
[1]Wu Z F, Dai C H, Yu J L, et al. Investigation of Blackbody 3 500 M[C]// SPIE. 2011 International Symposium on Phtoelectronic Detection and Imaging (ISPDI). Beijing, China,2011. v8192:819250-1-11.
[2]Woolliams E R,Harrison N J,Fox N P. Preliminary results of the investigation of a 3 500 K black body[J].Metrologia,2000, 37: 501-504.
[3]Sperfeld P,Metzdorf J,Harrison N J,et al.Investigation of high-temperature black body BB3200[J].Metrologia, 1998, 35: 419-422.
[4]WU Zhi-feng, DAI Cai-hong, YU Jia-Lin, et al. Realization of the 250 nm~400 nm spectral irradiance scale[C] // SPIE. 2011 International Conference on Optical Instruments and Technology: Optoelectronic Measurement Technology and Systems (OIT2011). Beijing, China, 2011.V8201:82012B-1-8.
[5]代彩红,于家琳,于靖,等. 颜色温度和相关色温的不确定度评定方法[J]. 光学学报,2005, 4(25): 547-552.
[6]Ouyang H Q,Dai C H,Huang B,et al.Effect of slit widths on wavelength calibration of a Czerny-Turner double monochromator[C]// Proceedings of SPIE. 2011 International Conference on Optical Instruments and Technology: Optoelectronic Measurement Technology and Systems (OIT2011). Beijing, China, 2011.v8201:820122-1-9.
[7]Ouyang H Q,Dai C H,Huang B,et al.Bandwidth determination of a Czerny-Turner double monochromator with varied slit widths through experiment and data computation[C] // Proceedings of SPIE. 2011 International Conference on Optical Instruments and Technology: Optoelectronic Measurement Technology and Systems (OIT2011). Beijing, China, 2011.v8197:81971B-1-9.
[8]Yoon H W,Proctor J E,Gibson C E. FASCAL 2: A New NIST Facility for the Calibration of the Spectral Irradiance of Sources[J].Metrologia,2003, 40(1): S30-S34.
[9]Friedrich R,J.Fischer New spectral radiance scale from 220 nm to 2 500 nm[J]. Metrologia, 2000, 37:539-542. |
|
|
|