Research on Optimal Amplitude of Coriolis Mass FlowmeterBased on Experimental Data Analysis |
HUANG Ya1,2,XU Kejun1,3,LIU Chenci1,2 |
1. School of Electrical and Automation Engineering, Hefei University of Technology, Hefei, Anhui 230009, China
2. Hefei Geyi Integrated Circuit Co., Ltd., Hefei, Anhui 230061, China
3. Engineering Technology ResearchCenter of Industrial Automation, Anhui Province, Hefei, Anhui 230009, China |
|
|
Abstract In response to the difficulty in determining the optimal amplitude and achieving the best measurement effect of the Coriolis mass flowmeter, the test was carried out by matching the developed transmitter with the primary instrument with excellent performance. The signals were collected under different amplitudes to analyze the impact of amplitude on the performance of the Coriolis mass flowmeter from the perspectives of amplitude stability, waveform distortion, signal power spectrum, signal-to-noise ratio, and system driving efficiency. The water flow calibration experiments were conducted with different amplitudes to determine the optimal amplitude based on indication error and repeatability. The results show that as the amplitude increases, waveform distortion becomes more severe, higher harmonic components also increase, and the noise signal power also shows an increasing trend, thereby affecting the signal-to-noise ratio and reducing the driving efficiency. However, if the amplitude is too small, it will exacerbate the amplitude fluctuation of the amplified signal of the sensor and affect the measurement results. Therefore, the optimal amplitude for this flowmeter is approximately 439mV.
|
Received: 12 June 2023
Published: 04 July 2024
|
|
|
|
|
[1] |
张启平, 魏祥毓, 唐秋林, 等. 科里奥利质量流量计的激振与信号提取 [J]. 仪器仪表与分析监测, 1992(2): 1-4.
|
[6] |
王利军. 基于振动管测量系统建模的科里奥利质量流量计零点补偿方法研究 [D]. 杭州: 浙江大学, 2013.
|
[7] |
乐静, 徐科军, 刘文, 等. 面向不同应用的两种科氏质量流量计信号处理方法 [J]. 电子测量与仪器学报, 2018, 32(5): 97-106.
|
[8] |
侯其立, 徐科军, 李叶, 等. 基于TMS320F28335的高精度科氏质量流量变送器研制 [J]. 仪器仪表学报, 2010, 31(12): 2788-2795.
|
|
ZHANG Q Q, WEI X Y, TANG Q L, et al. Excitation and Signal Extraction of Coriolis Mass Flowmeter [J]. Instrumentation and Analytical Monitoring, 1992(2): 1-4.
|
[15] |
李中翔, 郑德智, 胡纯, 等. 双管型科氏质量流量传感器非线性影响抑制研究 [J]. 仪器仪表学报, 2022, 43(1): 35-45.
|
[3] |
HENRY M P, CLARKE D W, ARCHER N, et al. A self-validating digital Coriolis mass-flow meter: an overview [J]. Control Engineering Practice, 2000, 8(5): 487-506.
|
[5] |
WANG L J, HU L, LEE K M, et al. Analytical Magnetic Field and Driving Force Models Based on Measured Boundary Conditions for Industrial Coriolis Mass Flowmeters [J]. IEEE Transactions on Industrial Electronics, 2012, 59(12): 4753-4760.
|
|
LE J, XU K J, LIU W, et al. Geared to the needs of different applications of two kinds of coriolis mass flowmeter signal processing method [J]. Journal of electronic measurement and instrument, 2018, 32 (5): 97-106.
|
[14] |
许爱强, 魏辉, 汪定国, 等. 基于Matlab估计正弦信号信噪比时频法研究 [J]. 测试技术学报, 2012, 26(1): 46-50.
|
|
TAO B B, XU K J, HOU Q L, et al. Control method for Coriolis mass flow tube with variable sensor settings [J]. Chinese Journal of Scientific Instrument, 2015, 36(3): 712-720.
|
[4] |
ZHENG D Z, WANG S, FAN S C. Nonlinear Vibration Characteristics of Coriolis Mass Flowmeter [J]. Chinese Journal of Aeronautics, 2009, 22(2).
|
|
HOU Q L, XU K J, LI Y, et al. Development of High Precision Coriolis Mass Flow Transmitter Based on TMS320F28335 [J]. Chinese Journal of Scientific Instrument, 2010, 31(12): 2788-2795.
|
|
HUANG Y, XU K J, LIU C C. Research on the influence of amplitude of Coriolis mass flowmeter on zero point [J]. Acta Metrology Sinica, 2023, 44 (2): 212-218.
|
[11] |
宋明刚, 樊尚春. 非线性对Coriolis质量流量测量管振动特性影响的理论研究 [J]. 仪器仪表学报, 2002, 23(4): 361-365.
|
[13] |
CHEESEWRIGHT R, CLARK C, BISSET D. Understanding the experimental response of Coriolis massflow meters to flow pulsations [J]. Flow Measurement & Instrumentation, 1999, 10(4): 207-215. DOI: 10.1016/S0955-5986(99)00014-X.
|
[16] |
段慧明, 史振东. 介绍2008版《科里奥利质量流量计》检定规程 [J]. 中国计量, 2009(4): 121-122.
|
[2] |
陶波波, 徐科军, 侯其立, 等. 变传感器设定值的科氏质量流量管控制方法 [J]. 仪器仪表学报, 2015, 36(3): 712-720.
|
[9] |
黄雅, 徐科军, 刘陈慈. 科氏质量流量计振幅对零点影响的研究 [J]. 计量学报, 2023, 44(2): 212-218.
|
|
LI Z X, ZHENG D Z, HU C, et al. Research on suppression of nonlinear influence of double-tube Coriolis mass flow sensor [J]. Instrument and Apparatus Journal, 2022, 43(1): 35-45.
|
[10] |
STORM R, KOLAHI K, ROCK H. Model-based correction of coriolis mass flowmeters [J]. IEEE Transactions on Instrumentation and Measurement, 2002, 51(4): 605-610.
|
|
SONG M G, FAN S C. Theoretical Research on Nonlinear Vibration Characteristics of Coriolis Mass Flow Measurement Tube [J]. Chinese Journal of Scientific Instrument, 2002, 23(4): 361-365.
|
[12] |
LI Z X, HU C, ZHENG D Z, et al. Synchronization Theory-Based Analysis of Coupled Vibrations of Dual-Tube Coriolis Mass Flowmeters [J]. Sensors, 2020,20(21): 6340.
|
|
XU A Q, WEI H, WANG D G, et al. Research on Time-frequency Method of Sinusoidal Signal to Noise Ratio Estimation Based on Matlab [J]. Journal of Measurement Technology, 2012, 26(1): 46-50.
|
[17] |
科里奥利质量流量计检定规程:JJG 1038-2008 [S]. 2008.
|
|
|
|