National Measurement Comparison for Verification Ability of Ophthalmometers |
ZHANG Ji-yan,YANG Lei,LIU Wen-li,GAO Ming-liang,CAO Yuan-hang |
National Institute of Metrology, Beijing 100029, China |
|
|
Abstract In order to investigate the accuracy and consistency of measurement values and issued measurement results for ophthalmometers by various laboratories, China National Institute of Metrology, as the pilot laboratory, organized and implemented the national Class A measurement comparison for radius verification ability of ophthalmometer from 2020 to 2021. A total of 33 laboratories participated in this measurement comparison, covering 15 provinces/municipalities directly under the Central Government, and the participating laboratories were highly representative. The comparison results were evaluated by the normalized error En value. With data processing and statistical analysis, the absolute value of En of 33 participated laboratories was between 0 and 0.85, all less than 1. The difference between the measurement results and reference values of each participated laboratory was within reasonable expectations. The results showed that the verification ability of radius of curvature in all participated laboratories met the technical requirements of measurement standards, and effectively ensured the accuracy, consistency and reliability of quantity values of ophthalmometers.
|
Received: 12 April 2022
Published: 21 September 2023
|
|
|
|
|
[2] |
李柯然, 李巧林, 徐向忠, 等. 眼轴长度与角膜曲率半径及其比值对儿童青少年屈光不正的影响和定性评估 [J]. 国际眼科杂志, 2019, 19(10): 1667-1671.
|
[3] |
王越, 柯敏, 王文欢, 等. 年龄相关性白内障患者眼轴长度与角膜曲率、角膜散光、前房深度和眼压的关系 [J]. 眼科新进展, 2017, 37(9): 879-882.
|
[1] |
李强强, 王悦, 郑康杰. 6~18岁儿童眼部屈光度、眼轴长度和角膜曲率分析 [J]. 预防医学, 2020, 32(9): 917-919.
|
[4] |
ISO 10343-2014 Ophthalmic instruments—Ophthalmometers [S]. 2014.
|
[5] |
张吉焱, 王莉茹, 陈燕. 角膜曲率计计量标准器的研制 [J]. 计量学报, 2006, 27(1): 28-31.
|
|
Zhang J Y, Hong B Y, Gao M L. Uncertainty evaluation for standard devices of ophthaomometers by using sample test method [J]. Measurement Technique, 2013(9): 47-50.
|
|
Zhang J Y, Wang L R, Chen Y. The development of standard devices for ophthalmometers [J]. Acta Metrologica Sinica, 2006, 27(1): 28-31.
|
|
Zhang J Y, Li J, Gao M L,et al . The development of axial standard device for calibration of ophthalmometers [J]. Acta Metrologica Sinica, 2017, 38(3): 276-279.
|
[9] |
陈燕, 洪宝玉, 张吉焱. 关于对具有角膜或接触镜曲率测量功能的验光仪的检定 [J]. 计量技术, 2014(12): 65-67.
|
|
Zhang J Y, Liu W L, Gao M L, et al. Calibration and traceability method for human cornea parameters [J]. Acta Metrologica Sinica, 2020, 41(11): 1339-1343.
|
[11] |
JJG 1011-2018 角膜曲率计 [S]. 2018.
|
[13] |
JJG 892-2022 验光仪 [S]. 2022.
|
[15] |
张以谟. 应用光学 [M]. 北京: 电子工业出版社, 2010.
|
|
Li Q Q, Wang Y, Zheng K J. Analysis of eye diopter, axial length and corneal curvature in children aged 6~18 years [J]. Journal of Preventive Medicine, 2020, 32(9): 917-919.
|
|
Li K R, Li Q L, Xu X Z, et al. Effect of axial length and corneal curvature radius and their ratio on refractive errors in children and adolescents [J]. Int Eye Sci, 2019, 19(10): 1667-1671.
|
|
Wang Y, K M, Wang W H, et al. Effects of axial length on corneal curvature, corneal astigmatism, anterior chamber depth and intraocular pressure in age-related cataract [J]. Recent Advances in Ophthalmology, 2017, 37(9): 879-882.
|
[7] |
张吉焱, 洪宝玉, 高明亮. 样板检验法测量角膜曲率计用计量标准器的不确定度评定 [J]. 计量技术, 2013(9): 47-50.
|
|
Chen Y, Hong B Y, Zhang J Y. Verification of eye refractometers with curvature measurement function of cornea or contact lens [J]. Measurement Technique, 2014(12): 65-67.
|
[14] |
JJF 1117-2010 计量比对技术规范 [S]. 2010.
|
[16] |
GB/T 26397-2011 眼科光学 术语 [S]. 2011.
|
[6] |
张吉焱, 李姣, 高明亮, 等. 角膜曲率计轴位标准器的研制 [J]. 计量学报, 2017, 38(3): 276-279.
|
[8] |
Zhang J Y, Liu W L, Gao M L, et al. Metrological calibration of ophthalmometers [C]//The 8th International Conference on BioMedcial Engineering and Informatics. Beijing, China, 2015:360-365.
|
[12] |
JJG 1088-2019 角膜曲率计用计量标准器 [S]. 2019.
|
[10] |
张吉焱, 刘文丽, 高明亮, 等. 人眼角膜参数计量校准与量值溯源方法研究 [J]. 计量学报, 2020, 41(11): 1339-1343.
|
|
|
|