Experimental Investigation of a Reciprocating Double-plunger Gas Flow Standard Device
XU Chun-wei1,ZHANG Gao-ming2,XU Zhi-peng2
1. Zhejiang Institute of Mechanical & Electrical Engineering, Hangzhou, Zhejiang 310053, China
2. Zhejiang Provincial Key Laboratory of Flow Measurement Technology, China Jiliang University,
Hangzhou, Zhejiang 310018, China
Abstract To solve the problems of restricted dimension of cylinder and limited flow range of the plunger gas flow standard devices, a reciprocating double-plunger gas flow standard device was developed, operating modes including single-cylinder, parallel double-cylinder and tandem double-cylinder modes were explored. For different operating modes, the repeatability and accuracy of the device were verified by the calibration experiment of a sonic nozzle with the nominal flow rate of 0.06m3/h. Results revealed that the repetitive rate of the device was less than 0.02% at different modes, the errors of the device at single-cylinder and parallel double-cylinder modes were less than 0.02%, while the error of the device at tandem double-cylinder mode was less than 0.05%.
XU Chun-wei,ZHANG Gao-ming,XU Zhi-peng. Experimental Investigation of a Reciprocating Double-plunger Gas Flow Standard Device[J]. Acta Metrologica Sinica, 2023, 44(8): 1229-1234.
Li M N, Li C H. Exploring the research direction of fluid flow measurement in China based on the distribution of calibration and measurement capabilities [J]. Acta Metrologica Sinica, 2022, 43(7): 885-891.
Ma J, Xu K J, Jiang Z, et al. A signal processing method of ultrasonic gas flowmeter based on peak point fitting of ultrasonic echo energy[J]. Acta Metrologica Sinica, 2022, 43(5): 597-602.
[7]
Wang C, Li X, Wang D W, et al. Establishment of piston devices[C]//FLOMEKO 12th, The MEKO Technical Committee on Flow Measurement. Guilin, China. 2004: 184-189.
[9]
Padden H. Uncertainty analysis of a high-speed dry piston flow prover [C]//FLOMEKO 12th. 2004.
[12]
Bellinga H, Delheza F J. Experience with a high-capacity piston prover as a primary standard for high-pressure gas flow measurement [J]. Flow measurement and Instrumentation, 1993, 4(2): 85-89.
Zhang Y S, Bai Y. The Passive Piston Gas Primary Standard[J]. Journal of Astronautic Metrology and Measurement, 2011(1): 67-70.
Zhao Y J, He G L, Xu Z. Numerical Simulation and Experiment Study on Resistance Characteristics of Coriolis Mass Flowmeter for 35MPa/70MPa Hydrogen Dispensers[J]. Acta Metrologica Sinica, 2022, 43(11): 1445-1449.
Feng J, Cheng J, Liang G W. Experimental study on plunger calibration devices of critical flow nozzles [J]. Journal of China University of Metrology, 2010, 21(2): 108-112.
Li P J, Chui L S, Li C H. The inner diameter measurement and uncertainty evaluation on the piston cylinder of piston gas flow standard device [J]. Acta Metrologica Sinica, 2021, 42(10): 1275-1281.
Lü C Z, Li M N, Li C H, et al. Influence of pipe diameter size deviation on the measurement performance of ultrasonic flowmeter [J/OL]. China Measurement & Test. http://kns.cnki.net/kcms/detail/51.1714.TB.20220811.1149.002.html.
[8]
Alasia F, Cignolo G. Design criteria for a 1200-liter capacity piston prover for primary gas volume and flowrate measurements [C]//FLOMEKO 4th, The MEKO Technical Committee on Flow Measurement. Australia. 1985.
[10]
Wright J D. What is the ‘Best’ Transfer Standard for Gas Flow? [C]//Proceedings of Flomeko. 2003.
[11]
Bellinga H. Using a piston prover as a primary standard in high-pressure gas mertering[C]//FLOMEKO 4th, The MEKO Technical Committee on Flow Measurement, Australia. 1985: 129-134.