Abstract The application of Van Der Pauw theory to solution electrical conductivity measurement is proposed innovatively, a closed four-electrode conductivity cell is designed. A micro AC current source is used to provide excitation signals,control circuit is used to process signals and communicate with computer.A series of experiments are carried out to study the influence factors of Van Der Pauw method.The tests of the reference materials proved the measurement range of the method is 0~40 mS/cm,the measurement error is under ±1%,which can be used for primary standard methods.
|
|
|
|
|
[1]宋小平.溶液电导率的绝对测量方法[J].化学分析计量, 2004,13(6): 88-89.
[2]宋小平. JONES型电导池测量的LCR电桥等效电路选择[J].化学分析计量, 2004,13(6): 79-82.
[3]Moroń Z, Rucki Z, Szczepanik Z.The possibility of employing a calculable four-electrode conductance cell to substitute for the secondary standards of electrolytic conductivity[J]. IEEE Transactions on Instrumentation and Measurement,1997,46(6): 1268-1273.
[4]Moroń Z. Investigations of van der Pauw method applied for measuring electrical conductivity of electrolyte solution[J]. Measurement, 2003,33(3): 281–290.
[5]Szczepanik Z, Rucki Z. Field analysis and electrical models of multi-electrode impedance sensors[J]. Sensors and Actuators A: Physical,2007,133(1):13–22.
[6]王洪伟, 张训时. 一种新型四电极式电导率探头的分析[J].仪器仪表学报, 1998,19(4): 399-402. |
|
|
|