Research on the Factors Influencing the Calibration of Medical Diagnostic Half-value Layer Instrument |
WEI Peng1,WANG Jia-wei2,LI De-hong3,MA Bo-xuan3,GUO Bin3 |
1.Fujian Institute of Metrology,Fuzhou,Fujian 350003,China
2.Anhui Institute of Metrology,Hefei, Anhui 230051,China
3.National Institute of Metrology,Beijing 100029,China |
|
|
Abstract The half-value layer measurement is required in various diagnostic X-ray radiation source verification regulations or calibration specifications, and half value layer measuring instruments are also widely used, but the calibration method of half value layer measuring instruments has not been clear. The main influencing factors in the calibration of the half value layer measuring instrument are discussed. The results show that the detector angle has the greatest influence on the measurement results. Therefore, the half value layer measuring instrument to be calibrated should be placed in the center of the test plane during calibration, and the angle deviation should not exceed ±1°.
|
Received: 08 February 2021
Published: 13 January 2023
|
|
|
|
|
[1] |
黎亚平, 吴丽萍, 谢万, 等. X射线质量吸收系数测量中一些问题的研究[J]. 四川大学学报(自然科学版), 2005, 42(2): 343-346.
|
[6] |
金玉博, 江南. 半价层测量方法的深入研究[J]. 中国医疗器械信息, 2014(3): 51-53.
|
[8] |
JJG1145—2017 医用乳腺X射线辐射源[S]. 2017.
|
[2] |
眭松山, 魏军, 史青. 质量衰减系数的计算与应用[J]. 光谱学与光谱分析, 1996, 16(2): 85-89.
|
|
Sui S S, Wei J, Shi Q. The Calcylation and Application of Mass Attenuation Conefficients[J]. Spectroscopy and Spectral Analysis, 1996, 16(2): 85-89.
|
|
Wang T Q, Wei X D, Li H J, et al. Calculation of X-ray Attenuation and Energy Deposition [J]. Atomic Energy Science and Technology, 2007, 41 (4): 399-403.
|
[5] |
王鹏, 邢立腾,蒋伟. 低空气比释动能率系列过滤X射线辐射质的测量与研究[J]. 计量学报, 2021, 42(4): 521-525.
|
[7] |
JJG744—2004 医用诊断X射线辐射源[S]. 2004.
|
[10] |
邹建新, 王鹏德, 李小双, 等. 医用诊断X射线半价层的测量[J]. 上海计量测试, 2020, 47(2):17-20.
|
|
Guo S M, Liu Y, Wu J J. Half Value Layer Measurement of Mammography X-Ray Standard. [J]. Acta Metrologica Sinica, 2018, 39(3): 410-413.
|
[13] |
JJF1621—2017诊断水平剂量计校准规范[S]. 2017.
|
[15] |
郭彬, 樊成. 诊断X射线辐射质RQR系列建立方法研究[J]. 计量学报, 2011, 32(6A): 45-47.
|
|
Guo B, Fan C. Investigation of the Diagnostic X-ray RQR Series Establishment[J]. Acta Metrologica Sinica, 2011, 32(6A): 45-47.
|
[18] |
YY/T 0722—2016 医用电气设备在诊断放射学中用于X射线管电压非接入式测量的剂量学仪器[S]. 2016.
|
|
Li Y P, Wu L P, Xie W, et al. The Study on the Measurement of X-Rays Mass Absorption Coefficients[J]. Journal of Sichuan University (Natural Science Edition), 2005, 42(2): 343-346.
|
[3] |
王同权, 魏晓东, 李宏杰, 等. X射线的衰减和能量沉积计算[J]. 原子能科学技术, 2007, 41 (4): 399-403.
|
[4] |
郝艳梅, 吕雅竹,李德红, 等. 热释光剂量计测量X射线半值层法初步探讨[J]. 计量学报, 2021, 42(6): 811-814.
|
|
Jin Y B, Jiang N. Further Research on HVL Measurement Methods[J]. China Medical Device Information, 2014(3): 51-53.
|
[9] |
JJG1078—2012 医用数字摄影(CR、DR)系统X射线辐射源[S]. 2012.
|
|
Zou J X, Wang P D, Li X S, et al. Measurement of half-value layer for medical diagnostic X-ray[J]. Shanghai Measurement and Testing, 2020, 47(2):17-20.
|
[14] |
YY/T0481—2016医用诊断X射线设备测定特性用辐射条件[S]. 2016.
|
[16] |
李兵, 吴金杰, 方方, 等. 医用诊断X射线束半值层和同质系数的研究与测量[J]. 核电子学与探测技术, 2012, 32 (8): 946-952.
|
[17] |
李德红, 郭彬, 黄建微, 等 . 诊断X射线辐射质的模拟与测量研究[J]. 中华放射医学与防护杂志, 2017, 37(8): 614-617.
|
|
Hao Y M, Lü Y Z, Li D H, et al. Initial Exploration of Half-value Layer Determination Using Thermoluminescent Dosimeter[J]. Acta Metrologica Sinica, 2021, 42(6): 811-814.
|
|
Wang P, Xing L T, Jiang W. Measurement and Research for the Filtered X-ray Radiation Quality of Low Air Kerma Rate Series[J]. Acta Metrologica Sinica, 2021, 42(4): 521-525.
|
[11] |
郭思明, 刘莹, 吴金杰. 乳腺X射线基准装置半值层的测量[J]. 计量学报, 2018, 39(3): 410-413.
|
[12] |
廖旭辉; 基于半导体探测器的乳腺钼靶机X射线检定装置的研究[D]. 衡阳:南华大学.
|
|
Li D H, Guo B, Huang J W, et al. Study on simulation and measurement of diagnostic X-ray radiation qualities [J]. Chin J Radiol Med Prot, 2017, 37(8): 614-617.
|
|
Li B, Wu J J, Fang F, et al. The Researchand Measurement of Medical Diagnostic X-ray Beam Half-Value Layer and Homogeneous Coefficient [J]. Nuclear Electronics & Detection Technology, 2012, 32 (8): 946-952.
|
|
|
|