|
|
Experimental Research on the Calorimetry Method and Radiation Force Balance Method for Measurement of Ultrasound Power |
HAN Ming-sheng 1,YANG Ping2,ZHU Hai-jiang1 |
1. College of Information Science & Technology, Beijing University of Chemical Technology, Beijing 100029, China;
2. National Institute of Metrology, Beijing 100013, China |
|
|
Abstract Based on the principle of the volume of the castor oil and the buoyance changed when the energy is absorbed by the castor oil, it has designed and achieved the target filled with the castor oil and suspended in the water. And further it has established the system of measurement of the ultrasound power using the balance for the displaying the change of the target, which is solved the traditional calorimetry to accurately measure the temperature field distribution problem. Compared to the plane wave power measured by the radial force balance method, the effectiveness of the measurement system using the calorimetry method for measurement of the acoustic power is validity. And further study of high power and non-plane wave sound field will be carried out.
|
|
|
|
|
[1]Dubinsky T J, Cuevas C, Dighe M K, et al. High-Intensity Focused Ultrasound: Current Potential and Oncologic Applications [J]. American Journal of Roentgenology,2008,190(1):191-199.
[2]Orsi F,Zhang L, Arnone P, et al. High-Intensity Focused Ultrasound Ablation: Effective and Safe Therapy for Solid Tumors in Difficult Locations[J]. American Journal of Roentgenology, 2010,195(3):W245-W252.
[3]黄小唯,寿文德. 毫瓦级超声功率测量系统的研制[J].应用声学,2004,23(1):29-34.
[4]顾欣. 一种基于量热法的超声功率测量方法的研究[D]. 上海:上海交通大学, 2006.
[5]Shaw A. A Buoyancy Method for the Measurement of Total Ultrasound Power Generated by HIFU Transducers [J].Ultrasound in Medicine and Biology,2008,34(8):1327-1342.
[6]Rajagopal S, Shaw A.The Buoyancy Method—a Potential New Primary Ultrasound Power Standard[J]. Metrologia,2012,49(3): 327-339. |
|
|
|