Spectral Responsivity Calibration of Infrared Radiometer Based on Piecewise Linear Calibration Method |
HE Ying-feng,DENG Le-wu,WU Jie,WEI Ping,WANG Rui |
Chengdu Aircraft Industrial (Group) Co. Ltd., Chengdu 610000, China |
|
|
Abstract The responsibility characteristics of the SR 5000N measurement system is calibrated based on blackbody radiation source which meets the national high temperature standard. A calculation model of the spectral responsivity calibration is established and the calibration is carried out by using the SR 5000N measurement system at 373~773K. The experiment results show that the measured infrared spectra in the wavelength range of 8 ~14μm at 373~773K have a low relative average deviation,which is less than 1%, compared with the calculation.
|
Received: 30 January 2019
Published: 08 June 2020
|
|
|
|
|
[1]董伟, 王卜平, 王景辉, 等. 一种高发射率涂料的红外发射率测量 [J]. 计量学报, 2012, 33(z2): 73-76.
Dong W, Wang B P , Wang J H,et al. Spectral Emissivity Measurement of a High Emissivity Black Coating Sample in the IR Wavelength Range [J]. Acta Metrologica Sinica, 2012, 33(z2): 73-76.
[2]薛生虎, 李文军, 李芳红, 等. 辐射测温在钢铁工业中的应用及发射率对测量的影响 [J]. 计量学报, 2010, 31(5): 445-449.
Xue S H, Li W J, Li F H,et al. Applications of Pyrometry in Rolling Mills and Study of the Effect of Emissivity on On-line Temperature Measurements of Steel Slab [J]. Acta Metrologica Sinica, 2010, 31(5): 445-449.
[3]白洁 刘绍民,扈光. 针对TM/ETM+遥感数据的地表温度反演与验证 [J]. 农业工程学报, 2008, 24(9): 148-154.
Bai J, Liu S M, Hu G. Inversion and verification of land surface temperature with remote sensing TM/ETM+ data [J]. Transactions of The Chinese Society of Agricultural Engineering, 2008, 24(9): 148-154.
[4]王军, 高教波, 胡煜. 无人侦察机红外辐射特性测试与视距估算 [J]. 红外与激光工程, 2008, (S2): 504-509.
Wang Jun, Gao J B, Hu Y. Infrared radiation characteristic measuring and visual range prediction of reconnaissance ummanned aerial vehicle [J]. Infrared and Laser Engineering, 2008, (S2): 504-509.
[5]Amditis A. Fusion of infrared vision and radar for estimating the lateral dynamics of obstacles [J]. Information Fusion, 2005,6(2): 129-141.
[6]Zhao X K, Zhao Q W, Wang L F. Laser and infrared compatible stealth from near to far infrared bands by doped photonic crystal[J]. Procedia Engineeringt, 2011,15: 1668-1672.
[7]张可佳,张碧丰,熊利民,等. EL红外缺陷测试仪关键参量校准方法研究[J]. 计量学报, 2018, 39(2): 246-250.
Zhang K J, Zhang B F, Xiong L M, et al. Research on Calibration Method Key Parameters of EL Infrared Defect Tester[J]. Acta Metrologica Sinica, 2018, 39(2): 246-250.
[8]薛程泓. 基于蒙皮的红外辐射能量测量方法研究[D]. 沈阳: 沈阳大学, 2015.
Xue C H. Based on the Skin of the Infrared Radiation Energy Measurement Method Research[D]. Shenyang: Shenyang University, 2015.
[9]任宁. 红外成像制导技术的发展 [J]. 红外与激光工程, 2007, 36(z2): 99-102.
Ren N. Development Of the infrared imaging guidance technology [J]. Infrared and Laser Engineering, 2007, 36(z2): 99-102.
[10]龚律宇, 郝小鹏, 孙建平, 等. H500型红外遥感定标高精度真空黑体辐射源的研制 [J]. 计量学报, 2017, 38(2): 129-134.
Gong L Y, He X P,Sun J P, et al. Development of H500 Type High Precision Vacuum Blackbody Used as a Calibration Standard for Infrared Remote Sensing [J]. Acta Metrologica Sinica, 2017, 38(2): 129-134. |
|
|
|