Investigation on the Calibration Methods of Gas Flow Standard Facilities by Master Method |
SONG Jin,LIU Yi-ping,CHEN Chao,LI Hai-yang |
Shanghai Institute of Measurement and Testing Technology, Shanghai 200231, China |
|
|
Abstract In view of the problems existing in the calibration of gas flow standard facilities by master method, such as short verification period , complicated removal and installation of master meter, a new method to trace the value of master meter is proposed and by experiments ,the feasibility of this method is verified. Experimental results show that: the meter factor get from the new method is no more than 0.1% different from the traditional method,and the error of the flowmeter calibrated by the flow standard facilities and higher level standard facilities shall not exceed 0.1%.
|
|
|
|
|
|
[1]王自和, 范砧. 气体流量标准装置[M]. 北京: 中国计量出版社, 2005, 312.
[2]郭爱华. 标准表法气体流量标准装置[J]. 自动化仪表, 2009, (30): 50-52.
[3]苏彦勋, 杨有涛. 流量检测技术[M]. 北京: 中国质检出版社, 2012.
[4]Tuskamoto H, Hutton S P. Theoretical prediction of meter factor for a helical turbine flowmeter[C]// Conference on Fluid Control and Measuement, Tokyo, Japan, 1986.
[5]段慧明, 史振东. 标准流量计的计量性能评定[J]. 计量技术, 2012, (S1): 62-65.
[6]国家质量监督检验检疫总局. JJG 643标准表法流量标准装置检定规程[S]. 2003.
[7]李春辉, 李鹏. 作为标准表非定点使用的涡轮流量计评定方法探析[J]. 计量学报, 2015, 36(6): 610-612. |
|
|
|