Research on Measuring Leakage Current ofDC High-voltage Resistive Divider |
ZHANG Huang-hui1,2,LIN Fei-peng3,SHAO Hai-ming3,ZHANG Zhong-hua1,3 |
1. School of Instrumentation Science and Opto-electronics Engineering, Beihang University, Beijing 100191, China
2. Fujian Metrology Institute, Fuzhou, Fujiang 350001, China
3. National Institute of Metrology, Beijing 100029, China |
|
|
Abstract A device using to measure the leakage current of DC high voltage resistor divider is introduced. The device uses ZigBee wireless communication technology, which by way of an independent power supply to avoid the risk of high voltage electric field resulting in damage to the measuring instrument, and solve issues of difficult to measure the leakage current of the voltage divider in the high voltage state. It can overcome the effect of high voltage power supply ripple and drift by using synchronous sampling method, and the uncertainty of measurement is less than 9.8×10-6 V, k=2.
|
Received: 22 February 2017
Published: 29 December 2017
|
|
|
|
|
[1]王乐仁, 章述汉. 电压加法在直流高压分压器电压系数测量中的应用[J]. 计量学报, 2003, 24(1):45-48.
[2]Li Y, Ediriweera M K, Emms F S, et al. Development of Precision DC High-Voltage Dividers[J]. IEEE Transactions on Instrumentation & Measurement,2011,60(7):2211-2216.
[3]Knight R B D, Martin P. A high voltage divider having an uncertainty of 5 ppm at 100 kV [J]. IEEE Transactions on Instrumentation & Measurement, 1993, 42(2): 568 - 570.
[4]Li Q, Wang L, Zhang S, et al. Method to Determine the Ratio Error of DC High-Voltage Dividers[J]. IEEE Transactions on Instrumentation & Measurement, 2012, 61(4):1072-1078.
[5]邵海明,李毅,林飞鹏.直流高电压标准的研究[R]. 中国计量科学研究院,2005.
[6]Marx R. New concept of PTBs standard divider for direct voltage of up to 100kV[J]. IEEE Transactions on Instrumentation & Measurement, 2001,50(2):426-429.
[7]李登云,刘高佳,李前,等.高电压下对精密电阻温度系数与电压系数的测量[J]. 电测与仪表,2012,49(10):20-24.
[8]Mccomb T R. A two-channel current measuring system for measuring leakage current in high direct volage measurements[C]// Conference on Precision Electromagnetic Measurements (CPEM). Korea, Daejeon, 2010:533-534.
[9]Lin F, Zhang H, Shao H, et al. Research on voltage variation of DC high-voltage resistive divider by Measuring Leakage Current[C]// Conference on Precision Electromagnetic Measurements(CPEM). Brazil, Rio de Janeiro, 2014:384-385.
|
|
|
|