Calibrating Pyrometers with High Temperature Fixed Points |
LU Xiao-feng1,YUAN Zun-dong1,DONG Wei1,WANG Tie-jun1,LUAN Hai-feng2,XIAO Xiao-ting3 |
1. National Institute of Metrology, Beijing 100029, China
2. Jiangsu Institute of Metrology, Nanjing, Jiangsu 210023, China
3. Institute for Metrology and Calibration of Guizhou, Guiyang, Guizhou 550003,China |
|
|
Abstract A 0.65 μm and a 0.9 μm pyrometers are interpolated by three high temperature fixed points (HTFPs), the best uncertainty is estimated 0.3~1.2 ℃ (k=2) in 900~2 500 ℃ for the 0.65 μm pyrometer and 0.3~0.6 ℃ in 600~2 000 ℃ for the 0.9 μm one separately. The pyrometers are also verified by the other fixed points and high temperature blackbody against the primary pyrometer, and the agreement is better than 0.1 ℃. The HTPFs can be applied in the precise calibration of pyrometers to improve the level of measurement uncertainty in high temperature range.
|
Received: 08 October 2016
Published: 11 August 2017
|
|
|
|
|
[1]卢小丰,原遵东,王景辉,等. 以“固定点-高温计”方式传递961.78 ℃以上国际温标[J].计量学报, 2014,35(3):193-197.
[2]卢小丰,柏成玉,王景辉,等. 工作基准钨带灯的新校准方法[J].计量学报,2012, 33(6A):58-60.
[3]JJG 1032—2007 标准光电高温计检定规程[S] .
[4]Yamada Y, Sakate H, Sakuma F,et al. Radiometric observation of melting and freezing plateaus for a series of metal-carbon eutectic points in the range 1330 ℃ to 1950 ℃[J].Metrologia,1999,36(3):207-209.
[5]Woolliams E R, Machin G, Lowe D H, et al. Metal (carbide)-carbon eutectics for thermometry and radiometry: a review of the first seven years[J].Metrologia,2006,43(6):R11–R25.
[6]Machin G. Twelve years of high temperature fixed point research: a review[C]// AIP Conference Proceedings for ITS2012, USA, Los Angeles,2013.
[7]Woolliams E R.,Anhalt K,Ballico M, et al. Thermodynamic temperature assignment to the point of inflection of the melting curve of high-temperature fixed points[J].Phil Trans R Soc A,2016,374(2064): 2015.0044.
[8]王铁军,柏成玉,原遵东,等.高温共晶固定点的建立与复现[J]. 计量学报,2007,27(3A):23-27.
[9]Dong W, Lowe D H, Lu X, et al. Bilateral ITS-90 Comparison at WC-C Peritectic Fixed Point Between NIM and NPL[C]// AIP Conference Proceedings for ITS2012, USA, Los Angeles,2013.
[10]Sakuma F, Hattori S. Establishing a Practical Temperature Standard by Using a Narrow-band Radiation Thermometer with a Silicon Detector [J]. Temperature :Its Measurement and Control in Science and Industry, 1982, 5: 421-427.
[11]王景辉.空腔黑体辐射源研制与评价[D].北京:清华大学,2015.
[12]Yuan Z D, Wang T J, Lu X F, et al. T90 Measurement of Co-C, Pt-C, and Re-C High-Temperature Fixed Points at the NIM[J].International Journal of Thermophysics,2011, 32(7):1744-1752.
[13]Saunders P, Fischer J,Sadli M, et al. Uncertainty Budgets for Calibration of Radiation Thermometers below the Silver Point[J]. International Journal of Thermophysics,2008,29(3):1066-1083. |
|
|
|