|
|
Value Expression and Uncertainty Evaluation of Relative Intensity Calibration Reference Material for Raman Spectroscopy |
YAO Ya-xuan,REN Ling-ling,GAO Si-tian,GAO Hui-fang |
National Institute of Metrology, Beijing 100029, China |
|
|
Abstract Raman spectroscopy is commonly used for molecular structure determination and qualitative and quantitative analysis. However, Raman spectra are rarely corrected for variations in spectrometer, which often leads to severe distortions in relative peak intensities, resulting in lack of comparability of Raman spectra between different instruments. Relative intensity calibration reference material and standard equipment of laser confocal microscopy Raman spectrometer is used, and homogeneity and stability of the reference material are examined. The results indicated that the standard reference material is homogeneous and the stated values are valid for one year, moreover, the correction curve of prepared reference material is consistent with that of NIST SRM2243.
|
Received: 21 December 2015
Published: 19 April 2017
|
|
|
|
|
[1]刘照军,梁会琴,王金艳. 拉曼强度中的分子结构信息 [J]. 光散射学报, 2009, 21(1): 1-6.
[2]Tan P H, Han W P, Zhao W J. The shear mode of multilayer graphene [J]. Nature Materials, 2012, 11(4): 294-300
[3]Hurst W S, Choquette S J, Etz E S. Requirements for relative intensity correction of Raman spectra obtained by column-summing charge-coupled device data [J]. Applied Spectroscopy, 2007, 61(7): 694-700.
[4]Fryling M, Frank C J, McCreery R L. Intensity calibration and sensitivity comparisons for CCD/Raman spectrometers [J]. Applied Spectroscopy, 1993, 47(12): 1965-1974.
[5]姚雅萱,任玲玲,高思田,等. 拉曼光谱仪相对强度校准及不确定度评定 [J]. 现代科学仪器, 2015,(3): 134-139.
[6]赵明媛,任玲玲,林智国,等. 拉曼光谱强度可控的硼硅锌光学玻璃制备[J]. 硅酸盐学报, 2012, 31(1): 258-261.
[7]全浩,韩永志. 标准物质及其应用技术 [M].第2版. 北京: 中国标准出版社,2003.
[8]张晓光,黄挺,张伟,等. 维生素K1冻干标准物质的研制 [J]. 计量学报,2015, 36(3): 333-336. |
|
|
|