|
|
Energy Calibration and Detection Efficiency Simulation of CdTe Detector |
YAO Xin-bo1,WU Jin-jie2,WANG Yu-long2,CHEN Cheng2,3,WANG Jia2,4,YANG Yang2 |
1. Zhejiang Institute of Metrology, Hangzhou, Zhejiang 310013, China
2. National Institute of Metrology,
Beijing 100029, China
3. Chengdu University of Technology, Chengdu, Sichuan 610059, China;
4. University of South China, Hengyang, Hunan 421000, China |
|
|
Abstract CdTe detector need to be scaled before it detects the single parallel photon source. Based on the energy calibration graph and energy resolution by standard point source, accurate internal structure by CT scanning and the results of physical model calculated by MCNP5 Monte Carlo code can be achived. The detector efficiency graph can be obtained on 1keV step from 20keV to 150keV energy range. The detection efficiency is very well at lower energy range and becomes decline at the energy of 27keV and 32keV due to outcome of X-ray escape summits. Then the detector efficiency increases and then decreases with the growth of energy.
|
Received: 27 May 2015
Published: 19 April 2017
|
|
|
|
|
[1] Ewa I O B, Bodizs D,Czifrus S, et al. Monte Carlo determination of full energy peak efficiency for a HPGe detector[J]. Applied Radiation and Isotopes,2001,55(1):103-108.
[2]Pavel Dryak,Petr Kovar. Experimental and MC determination of HPGe detector efficiency in the 40-2754 keV energy range for measuring point source geometry with the source-to-detector distance of 25 cm[J]. Applied Radiation and Isotopes,2006,64(10-11):1346-1349.
[3]张建芳, 赵广义, 王玉德,等.高纯锗探测器探测效率的MCNP模拟[J].吉林大学学报,2010,48(5):843-846.
[4]段小娟, 吴金杰, 杨元第,等.EGSnrc程序模拟低能X射线光管参数对射线能谱的影响[J].计量学报,2011,32(z1):51-53.
[5]卫晓峰, 刘立业, 肖运实,等.蒙特卡罗方法计算高纯锗探测器的全能峰效率[J].辐射防护,2013,33(3):158-163.
[6]黄勇, 杨磊, 杨汝基,等.碲化镉(CdTe)探测器的原理及医学应用[J]. 上海生物医学工程,2005,(4):221-225.
[7]王增勇, 汤光平, 李建文,等. 工业CT技术进展及应用[J]. 无损检测,2010,(7):504-508.
[8]李福龙, 张雄杰, 王仁波. MCNP中高斯展宽系数的确定方法与验证[J]. 核电子学与探测技术,2013, 33(10):1266-1270.
[9]余少杰, 郝锐, 张静. MCNP模拟HP-Ge探测器效率刻度曲线[J]. 核电子学与探测技术,2014, 34(1):14-16. |
|
|
|