The Evaluation of Uncertainty of Flow Totalizer Standard Facility |
SHEN Chao,PEI Quan-bin,LIU Bo-tao,XIA Bao-ding,ZOU Wei |
Sinopec Natural Gas Company Research Center, Jinan, Shandong 250000, China |
|
|
Abstract According to the JJF 1059《 Evaluation and Expression of Uncertainty in Measurement》, and combining with flow totalizer standard facility,with Daniel S600 + flow totalizer for turbine flowmeter in natural gas metrology as an example, the uncertainty analysis of standard equipment was made with GUM. According to the traceability requirements, the meters whose uncertainty is no better than 0.14% can be calibrated by GUM. In accordance with the traceability requirements, it can verificate /calibrate the flow totalizer which is no better than level 0.5.
|
Received: 19 October 2015
Published: 19 April 2017
|
|
|
|
|
[1]国家质量监督检验检疫总局.JJG 1003—2005 流量积算仪检定规程[S].2005.
[2]国家质量监督检验检疫总局.JJF 1059—2012 测量不确定度评定与表示[S].2012.
[3]闫文灿,肖迪,徐明,等.移动式天然气计量标准装置不确定度评定[J].计量学报,2014,35(1):54-60. |
|
|
|