Abstract:DC high voltage dividers was usually calibrated by the method of resistance or leakage current mesurement. The 100kV and below dividers voltage ratio was measured by resistance measurement,as leakage current could be ignored because of its little order of magnitude.And the leakage current measurement method was used in the dividers with resistance-measurement-layer-only structure.Voltage summation was applied in DC high voltage dividers measurement. The improved voltage summation mathematical model was built based on the new low-voltage side resistance out-put technology.An experiment was carried out on a 100kV DC divider, and normalized deviation method was used in the voltage summation experiment result evaluation. The experiment shows that the component uncertainty of improved voltage summation is 10-5~10-6, and the 100kV DC divider uncetainty is 1.3×10-5 (k=2).
刘方明,刘浩,李登云,肖凯,杜百稳. 基于改进电压加法的直流高压分压器测量方法研究[J]. 计量学报, 2024, 45(11): 1593-1600.
LIU Fangming,LIU Hao,LI Dengyun,XIAO Kai,DU Baiwen. Research on the Measurement of DC High Voltage Dividers Based on Improved Voltage Summation. Acta Metrologica Sinica, 2024, 45(11): 1593-1600.
CHILDERS C B, DZIUBA R F, LEE L H. A resistive ratio standard for measuring direct voltages to 10kV [J]. IEEE Transactions on Instrumentation and Measurement, 1976, IM-25(4): 505-508.
WANG L R, ZHANG S H. Voltage Summation Method for Determing the Voltage Dependence Ratio Variation of Precise High DC Voltage Dividers [J]. Acta Metrologica Sinica, 2003, 24(1): 45-48.
MAO Y, LI Q F, LI Y, et al. Intercomparison and Uncertainty Evaluation of 1000kV DC High-Voltage Divider [J]. Power System Technology 2012, 36(7): 38-42.
[2]
HIRAYAMA H, KOBAYASHI M, MURAKAMI K, et al. 10kV High-accuracy DC voltage divider [J]. IEEE Transactions on Instrumentation and Measurement, 1974, 23(4): 314-317.
LI D Y, LIU G J, LI Q, et al. Measurement with High Voltage for Temperature Coefficient and Voltage Coefficient of High Precision Resistors [J]. Electrical Measurement & Instrumentation, 2012, 49(10): 20-24.
WANG C, ZHANG L L, ZHANG G S. Research on Nonlinear Mechanish of DC Resistive Voltage Divider[J]. Acta Metrologica Sinica, 2022, 43(4): 521-525.
SHI Z M, ZHANG J T, PAN X L, et al. Self-calibration of the Phase Angle Errors of RVDs with Serial-parallel Connection Structure [J]. Acta Metrologica Sinica, 2020, 41(3): 301-305.
[19]
LI Q, WANG L, ZHANG S, et al. Method to Determine the ratio Error of DC High-Voltage Dividers [J]. IEEE Transactions on Instrumentation & Measurement, 2012, 61( 4) : 1072-1078.
[20]
直流电压比例标准装置自校准方法: JJF 1872-2020 [S].
LEI M, JIANG C Y, LI D Y. Review of High Voltage Ratio Standard Devices Technology [J]. High Voltage Engeering, 2021, 47(6), 1893-1904.
[5]
MARX R. New concept of PTBs standard divider for direct voltages of up to 100kV [J]. IEEE Transactions on Instrumentation and Measurement, 2001, 50(2): 426-429.
ZHANG H H, LIN F P, SHAO H M. Research on Measuring Leakage Current ofDC High-voltage Resistive Divider[J]. Acta Metrologica Sinica, 2018, 39(1): 83-88.
KNIGHT R B D, MARTIN P. A high voltage divider having an uncertainty of 5 ppm. at 100kV [J]. IEEE Transactions on Instrumentation and Measurement, 1993, 42(2): 568-570.
[6]
LI Y, RUNGIS J, KIM K T, et al. Interlaboratory comparison of high direct voltage resistor dividers [J]. IEEE Transactions on Instrumentation and Measurement, 1999, 48(2): 158-161.
ZHANG S H, WANG L R. Calibration Method for High DC Voltage Dividers Based on Voltage Summation [J]. Acta Metrologica Sinica, 2006, 32(11): 53-56.
WANG L R. Voltage Summation Method for Voltage Ratio Standards at Power Frequency [J]. Acta Metrologica Sinica, 1992, 13(3): 221-224.
LIU F M, XIAO K, WANG H Y, et al. The influence of Digital Multimeter Input Impedance in DC High Voltage Dividers Measurement [J]. Acta Metrologica Sinica, 2023, 44(5): 771-776.
LIU K P, LI D Y, ZHU K, et al. Traceability of 1kV DC Voltage Ratio Based on 2/1 Self-calibration Method [J]. High Voltage Engeering, 2022, 48(6): 2226-2233.