Research on Calibration Technology of ReciprocalMagnetoelectric Vibration Sensor
LIAN Xiangkun1,CAI Chenguang2,KONG Ming1,LIU Zhihua2,LÜ Siyao1
1. Collage of Metrology & Measurement Engineering, China Jiliang University, Hangzhou, zhejiang 310018, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:A new reciprocity calibration method for passive servo vibration sensors is proposed. This method is based on the basic principle of reciprocity calibration. According to the dual coil characteristics of passive servo vibration sensors, they are divided into calibrated sensors and reciprocal sensors. By establishing the vibration differential equation, the frequency response characteristics of the sensitivity ratio, sensitivity product, and reciprocity calibration coefficient of the two sensors are analyzed, and the corresponding calibration method is finally proposed. Based on this method, two experimental measurements are carried out to obtain the sensitivity ratio and sensitivity product, and combined with the reciprocal calibration coefficients, the calibration of passive servo vibration sensors is realized. And the experimental comparison with the laser interferometric method is carried out, and the experimental results show that the relative error of the calibration results of the two sensors is no more than 1% at the frequency of 1~100Hz.
WANG Z, CAI C G, YANG M, et al. Vibration Calibration Method Based on the Decoding of Heterodyne Interferometer Output S/PDIF Digital Signal [J]. Acta Metrologica Sinica, 2021, 42(3): 282-286.
TANG B, HE W, LIU A D. Research Progress on Broadband Electromagnetic Angular Vibration Calibration System [J]. Acta Metrologica Sinica, 2021, 42(10): 1328-1334.
[8]
Methods for the calibration of vibration and shock transducers-Part 21: Vibration calibration by comparison to a reference transducermethod:ISO16063-21[S]. 2003.
Methods for the calibration of vibration and shock transducers-Part12: Primary vibration calibration by the reciprocity method:ISO16063-12[S]. 2002.
CHI Q L, YANG X S, SHANG S K. Passive servo feedback multi-output low-frequency vibration sensor [J]. Chinese Journal Of Scientific Instrument, 2017, 38(1): 105-111.
CHEN Y, CHEN H J, FANG C, et al. Comparative Study on Constant Current Calibration Technology and Constant Voltage Calibration Technology of Vibration Pickup [J]. Metrology & Measurement Technique, 2021, 48(1): 24-26.
WANG B Y, ZENG F Z, FU C Y, et al. Study on In-situ Calibration of Precision Platinum Resistance Thermometer Based on Micro Gallium Fixed-point [J]. Acta Metrologica Sinica, 2022, 43(2): 210-214.
DONG X M, WANG M L. Rectification Error Calibration of Accelerometer Based on Centrifugal-Vibration Composite Device [J]. Acta Metrologica Sinica, 2022, 43(9): 1166-1171.
YANG M, ZHAO H X, LIU Z H, et al. Calibration of Low Frequency Accelerometer Based on the Earth′s Graviation Method [J]. Acta Metrologica Sinica, 2022, 43(8): 1064-1069.
WANG J H, ZUO A B, PENG Y X. Primary Vibration Calibration of High Temperature Piezoelectric Acceleration Sensor [J]. Acta Metrologica Sinica, 2022, 43(12): 1634-1638.
LIN J. Two Methods for the Reciprocity Calibration of Piezoelectric Accelerometers [J]. Acta Acustica, 1979(4): 309-314.
WU H X, HAN X L. The Method of Mutual Calibration Between Speed Sensorsand Acceleration Pick-ups [J]. Journal of Vibration, Measurement & Diagnosis, 1995(1): 23-27.
YANG M, CAI C G, LIU Z H, et al. Calibration method of low frequency vibration sensor based on guideway bending correction of long stroke shaker [J]. Journal of Vibration and Shock, 2022, 41(1): 116-120.
[9]
方翠. 磁电式振动传感器恒流激励测量技术研究 [D]. 杭州: 中国计量大学, 2020.
WU H X, HAN X L. A New Method for Reciprocal Calibration of Accelerometers [J]. Journal of Vibration, Measurement & Diagnosis, 2001, 21(2): 55-58.
L Q, WANG J B, ZHANG Y, et al. Design Study on the Measurability of Low-Frequency Standard Shaker [J]. Metrology Science And Technology, 2022, 66(4): 101-107.